Catalog Number 3501420
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 04/01/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: left side deflation.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) caucasian female patient underwent revision breast augmentation with a 175cc mentor smooth round spectrum.The patient noticed a change of fullness of the left breast.Left breast deflation was confirmed via mammogram performed on (b)(6) 2019.As a result, the device was explanted on (b)(6) 2019.
 
Additional Manufacturer Narrative
On (b)(6) 2019, mentor became aware that the replacement device used is: catalog number: 3501420 serial number: (b)(4).Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Device evaluation summary: the analysis results showed a crease in the anterior aspect.Leak testing revealed there was leakage within the crease measuring approximately less than 0.1 cm.No additional anomalies were observed.These kind of findings are consistent with a crease/fold failure which is a known inherent risk associated with the use of saline-filled mammary prostheses and may be the result of one or more of the following contributing factors: underinflation or overinflation of the device, continuous and sustained stresses to the device - such as too small a breast pocket and folding or wrinkling of the shell in the breast pocket.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.It should be noted that as part of our quality process, each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Based on the information reported and/or the product investigation, there is no evidence that the issue is related with the manufacturing process.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SMOOTH ROUND SPECTRUM
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key8708408
Report Number1645337-2019-14036
Device Sequence Number11223852
Product Code FWM
UDI-Device Identifier00081317001966
UDI-Public00081317001966
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/14/2018
Device Catalogue Number3501420
Device Lot Number6882516
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2019
Initial Date Received by Manufacturer 05/23/2019
Supplement Date Received by Manufacturer06/25/2019
07/22/2019
Initial Report FDA Received Date06/18/2019
Supplement Report FDA Received Date07/18/2019
07/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
-
-