ENCORE MEDICAL L.P. 3DKNEE; INSERT, 3D SIZE8 X 9MM R DCM Back to Search Results
Model Number 392-09-608
Health Effect - Clinical Code No Information (3190)
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to poly wear.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to poly wear.The time between the previous surgery and the surgery detailed in this investigation is unknown.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to poly wear.There were no findings during this investigation that indicate that the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Tthere are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3DKNEE
Common Device Name
INSERT, 3D SIZE8 X 9MM R DCM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8708536
Report Number1644408-2019-00560
Device Sequence Number2072930
Product Code JWH
UDI-Device Identifier00888912120197
UDI-Public(01)00888912120197
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K020114
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/28/2010
Device Model Number392-09-608
Device Catalogue Number392-09-608
Device Lot Number462542
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/20/2019
Supplement Date Received by Manufacturer06/25/2019
Initial Report FDA Received Date06/18/2019
Supplement Report FDA Received Date07/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-