STRYKER GMBH UNKNOWN APEX PIN IMPLANT Back to Search Results
Catalog Number UNK_SEL
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Nerve Damage (1979); Numbness (2415)
Event Date 11/01/2012
Event Type  Injury  
Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided. This complaint has been reported during a literature review performed by the post market surveillance group. No product identification is possible. More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event. The reviewed ifu, however, states that the deep venous thrombosis is one of the most frequent adverse effects involving the use of internal fracture fixation devices and adverse results may be clinically related rather than device related. Also, the surgeon must warn patients of surgical risks, and make them aware of possible adverse effects. The patient should be warned that the device cannot and does not replicate a normal healthy bone, that the device can break or become damaged as a result of strenuous activity or trauma. Based on the available information, no relation could be established between the device and the reported event. A review of the labeling did not indicate any abnormalities. If any further information is provided, the investigation report will be updated. Device disposition unknown.
 
Event Description
The manufacturer became aware of a post market clinical follow-up (pmcf) from (b)(6). The title of this report is ¿a post-market clinical follow-up (pmcf) of the treatment of fractures with the apex pins in conjunction with stryker external fixation systems¿ which was published in april 2019 and is associated with the stryker apex pin. Within that publication, post-operative complications/ adverse events were reported, which occurred between november 2012 to september 2018. It was not possible to ascertain specific device or patient information from the article, a review of the complaint handling database, however, revealed that the events have not been reported by the hospital or by the author of the publication, therefore 52 complaints were initiated retrospectively for different adverse events mentioned in the journal. This product inquiry addresses nerve lesion. The report states, ¿subject a2-1 was a (b)(6) year old (b)(6) male who had a fall from standing injuring his left elbow (radial head fracture and elbow dislocation). Orif of the left elbow with radial head arthroplasty was performed and a stryker external fixator with four apex pins was placed for temporary stabilization. His history was significant for current tobacco use. Discharged on the same day as his surgery and placed non-weightbearing status. Subject was seen 10 days post op in the clinic and stated he developed an inability to use his left wrist, hand and fingers two to three days after surgery. Onset was gradual but complete. He also complained of numbness in the hand along the dorsal aspect of the thumb. Examination showed complete¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN APEX PIN
Type of DeviceIMPLANT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH 2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH 2545
Manufacturer Contact
rose haas
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key8709371
MDR Text Key148404918
Report Number0008031020-2019-00639
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Type of Report Initial
Report Date 06/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Catalogue NumberUNK_SEL
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/22/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 06/18/2019 Patient Sequence Number: 1
-
-