Analysis of historical records: orthofix (b)(4) checked the internal records related to the controls made on the lag screw code 99-t93700, batch b1277461 (lot laser marked on the component is b1260852) before the market release.No anomalies have been found.The original lot, manufactured in 2019, was comprised of (b)(4) devices.All of them have already been distributed to the market.According to orthofix (b)(4) historical records, no other notifications have been received from this specific device lot.(mfr report 9680825-2019-00040).Orthofix (b)(4) checked the internal records related to the controls made on the lag screw code 99-t93705 batch b1277606 (lot laser marked on the component is b1275439) before the market release.No anomalies have been found.The original lot, manufactured in 2019, was comprised of (b)(4) devices.All of them have already been distributed to the market.According to orthofix (b)(4) historical records, no other notifications have been received from this specific device lot.(mfr report 9680825-2019-00041).Technical evaluation: the returned screws, received on may 23, 2019, were examined by orthofix (b)(4) quality engineering department.The returned screws were subjected to visual and dimensional check as per orthofix specification.Device code 99-t93700 (mfr report 9680825-2019-00040): the visual check evidenced presence of damaging signs on the device surface.The visual check also evidenced that the teeth on the surface of the lag ferrule were damaged.It was also evidenced that the teeth on the surface of the lag screw do not present contact signs that should be when the screw is locked to the nail, suggesting that the screw was not properly inserted in the nail.The dimensional check, performed where possible, did not evidence any anomalies.A functional check was not possible as the device is broken.Device code 99-t93705 (mfr report 9680825-2019-00041): the visual check evidenced presence of damaging signs on the device surface.The visual check also evidenced that the teeth on the surface of the lag ferrule were damaged.It was also evidenced that the teeth on the surface of the lag screw do not present contact signs that should be when the screw is locked to the nail, suggesting that the screw was not properly inserted in the nail.The dimensional check, performed where possible, did not evidence any anomalies.A functional check was not possible as the device is broken.Medical evaluation the information made available on the case together with the results of the technical evaluation were sent to our medical evaluator.Please find below an extract of the medical evaluation performed.May 28, 2019: "in this case, i noted that a piece of the locking mechanism of the chosen lag screw broke off during a second attempt to insert it.A duplicate size of lag screw was not available, so the surgeon used a shorter screw to complete the operation.The comment was that this was not ideal but was all that was available.The reason that the shorter lag screw was used was because the hospital did not have a second screw of the chosen length.We need to see the x-rays of the fixation to assess the final result".On june 8, 2019 with the results of the technical evaluation: "the original complaint suggested that the surgeon had attempted to insert one screw, but this failed and a piece of the screw broke off, so a second shorter screw was inserted.I now note that 2 screws were the subject of the original complaint: 99-t93700 and 99-t93705; the screw that was used in the end was 99-t93790.It is not clear which lag screw was the original one chosen, but logically it should be the 105 mm screw.The surgeon then tried the 100 mm screw, and then when this failed the 90 mm screw which was successful.There is no mention of the 95 mm screw, which would have been the logical next choice.This suggests that the hospital did not have a complete stock of lag screws, and it seems that they only had one of each size.The technical report makes the observation that the teeth at the proximal ends of the two screws have not been deformed in any way, whereas the teeth on the lag ferrule have been damaged in both cases.This suggests that the screw has come into contact with something before it is fully inserted, rather than contacting the nail which causes the screw to lock in place.The technical report is suggesting that the cause in this case is a failure to insert the perforator 193970 far enough, with the result that there was a collar of narrow cortical bone which caused the breakage of the teeth on the ferrules.This was we are suggesting the most likely cause of the events, but this is an interpretation of the facts as we have them.If the original choice of screw length was 105 mm and the final used screw was 90 mm, it is likely that this fixation was less than optimal.However, it may still serve the purpose if other parameters are favourable (fracture reduction, degree of comminution, position of screw in neck - central or offset, bone quality).As suggested before, we need to see some x-rays to make any comment on the final stability of this fixation".Final comments: the results of the technical evaluation concluded that the returned lag screws were originally conforming to orthofix specification.It is most likely that both lag screws broke due to the particular conditions of use (reaming procedure not properly performed).During the reaming procedure, which is aimed to prepare a bone hole suitable for screw insertion, perforator code 193970 should be inserted until its end position (with the blue tube against the targeting handle 193100).It is very likely in this event that this position was not reached so a hole suitable for the insertion of the larger part of the screw was not obtained.Because of hole dimension smaller than required one, the toothed crown of the screws would have gone into interference with bone lateral cortex, leading to the teeth opening and to screw locking mechanism activation before reaching the correct screw insertion depth (i.E.Teeth in the nail hole).The medical evaluation evidenced as follows: may 28, 2019: "in this case i note that a piece of the locking mechanism of the chosen lag screw broke off during a second attempt to insert it.A duplicate size of lag screw was not available, so the surgeon used a shorter screw to complete the operation.The comment was that this was not ideal but was all that was available.The reason that the shorter lag screw was used was because the hospital did not have a second screw of the chosen length.We need to see the x-rays of the fixation to assess the final result".On june 8, 2019 with the results of the technical evaluation: "the original complaint suggested that the surgeon had attempted to insert one screw, but this failed and a piece of the screw broke off, so a second shorter screw was inserted.I now note that 2 screws were the subject of the original complaint: 99-t93700 and 99-t93705; the screw that was used in the end was 99-t93790.It is not clear which lag screw was the original one chosen, but logically it should be the 105 mm screw.The surgeon then tried the 100 mm screw, and then when this failed the 90 mm screw which was successful.There is no mention of the 95 mm screw, which would have been the logical next choice.This suggests that the hospital did not have a complete stock of lag screws, and it seems that they only had one of each size.The technical report makes the observation that the teeth at the proximal ends of the two screws have not been deformed in any way, whereas the teeth on the lag ferrule have been damaged in both cases.This suggests that the screw has come into contact with something before it is fully inserted, rather than contacting the nail which causes the screw to lock in place.The technical report is suggesting that the cause in this case is a failure to insert the perforator 193970 far enough, with the result that there was a collar of narrow cortical bone which caused the breakage of the teeth on the ferrules.This was we are suggesting the most likely cause of the events, but this is an interpretation of the facts as we have them.If the original choice of screw length was 105 mm and the final used screw was 90 mm, it is likely that this fixation was less than optimal.However, it may still serve the purpose if other parameters are favourable (fracture reduction, degree of comminution, position of screw in neck - central or offset, bone quality).As suggested before, we need to see some x-rays to make any comment on the final stability of this fixation".Based on the results of the technical evaluation and on the evidences deriving from the medical evaluation, orthofix (b)(4) can conclude that these screws failed because of the particular conditions of use.Orthofix (b)(4) requested the x-rays of the case and the information about patient's health condition.Unfortunately, this information was not made available.Orthofix (b)(4) historical records shows that no other notifications have been received in regards to these specific devices lot.Orthofix (b)(4) continues monitoring the devices on the market.Please also kindly refer to mfr report 9680825-2019-00041.
|