| Model Number |
MMT-722NAB |
| Medical Device Problem Codes |
Mechanical Problem (1384); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had motor error.The customer¿s blood glucose level was unknown.The customer stated that reservoir lip ring was not damaged.The customer was declined troubleshooting for motor error.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device alarmed motor error during the basic occlusion test due to loose or protruded drive support disk.Unable to perform basic occlusion test, occlusion test, prime or a33 test, excessive no delivery test, displacement test due to motor error alarm.However, device passed rewind test and displacement test.Device had broken battery tube threads.
|
| |
|