| Catalog Number |
180403 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Pain (1994); Injury (2348)
|
| Date of Event |
05/24/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's left knee (patellofemoral component) was revised after patient complaint of pain, 7 months post robotic procedure.Patient was revised to a stryker total knee.The rep reported that no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding pain involving a mako femoral component was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Clinician review: no medical records were received for review with a clinical consultant.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's left knee (patellofemoral component) was revised after patient complaint of pain, 7 months post robotic procedure.Patient was revised to a stryker total knee.The rep reported that no further information is available.
|
| |
|