• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LEGION PS NP FEM SZ 7 LT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LEGION PS NP FEM SZ 7 LT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Catalog Number 71423227
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 05/27/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that prosthesis was removed due to infection being cleared up in a second stage revision surgery.First stage surgery was purely to treat infection and no components where removed until second stage.
 
Additional Manufacturer Narrative
The associated legion ps non porous femoral component and genesis ii ps poly tibial base were not returned for evaluation.Therefore a product analysis could not be performed.However, device details were provided.Consequently, the manufacturing records and complaint history review was conducted.The review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.The review of the complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.The related devices were sterilized according to sterilization release documentation from quality control.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.There is no information that would suggest the implanted devices failed to meet specifications.A relationship, if any, between the devices and the reported incident or adverse event could not be corroborated.A clinical evaluation noted that no relevant supporting documentation was provided, therefore a thorough medical investigation could not be performed.Without the return of the actual products involved, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however smith and nephew will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION PS NP FEM SZ 7 LT
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8712941
Report Number1020279-2019-02337
Device Sequence Number16706052
Product Code JWH
UDI-Device Identifier03596010611505
UDI-Public03596010611505
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K060742
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/05/2025
Device Catalogue Number71423227
Device Lot Number15MM05698
Initial Date Received by Manufacturer 05/27/2019
Supplement Date Received by Manufacturer05/27/2019
Initial Report FDA Received Date06/19/2019
Supplement Report FDA Received Date09/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-