| Catalog Number |
109.327 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Codes |
Pain (1994); Tissue Damage (2104)
|
| Date of Event |
05/11/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.Moreover,it was seen that the dentist has used more drills than recommended to perform the implant installation.The investigation of the complaint was carried out considering the analysis of the information given by the dentist in the guarantee form, visual conference of the product returned by the dentist and the evaluation of relevant production records.
|
| |
|
Event or Problem Description
|
|
(b)(4).The dentist reported that 1 month after the dental implant was installed in patient¿s mouth, its non-osseointegration was observed.Moreover, 42n.Cm of primary stability was achieved, the patient presented bone type ii and bone graft was placed.
|
| |
|
Search Alerts/Recalls
|