|
Device evaluation : one uniperc sample was received for evaluation in used condition and without its original packaging.Upon immediate visual inspection, it was found that the cuff burst.However, this defect happened during decontamination by smiths medical and will be excluded from investigation.It was also noted that the inflation lumen on tracheostomy tube is punctured in place where inflation line is connected with lumen.This issue can occur when the adjustable flange is over pulled above the last marking on the tube (close to 15mm connector), which is in conflict with instruction for use.The device instruction describes the following: "using the "initial flange settings" table above and the skin to anterior tracheal wall thickness estimated at stage 11.4, as a guide, adjust the tracheostomy tube flange (patient side) using the 10mm graduations on the tracheostomy tube to its correct length and lock the flange firmly in position by pushing the blue locking arm downwards until it is flush with the white flange hub." additionally, during manufacturing process, the devices are 100% inflation tested, which includes inflating each device cuff and leaving for a 12 hour period.Reductions in pressure over this time are considered a failure and the device would be rejected.Furthermore, the device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.Based on the evidence and testing, the complaint was confirmed.User interface was noted as the problem source as the issue was observed after placement.The most probable cause of the reported failure occurred during tracheostomy procedure due to contact with sharp edges, also in conflict with instruction for use to "guard against cuff damage by avoiding contact with sharp edges.".
|