|
An ovation ix abdominal stent graft system was implanted to treat an abdominal aortic aneurysm.Post polymer fill of the aortic body stent graft, a polymer leak was suspected.The physician elected to implant a palmaz stent graft and balloon at the level of the sealing rings to resolve a type ia endoleak, however the endoleak remained.The physician elected to conclude the procedure.The patient will continue to be monitored.
|
|
A clinical assessment of the reported event could not be completed due to a lack of receipt of relevant medical records and/or imaging.Several attempts were made to obtain medical records and medical images, but no response was received.As such, procedure related harms, event determination, off label conditions, related patient harms and patient disposition could not be assessed with medical records or images.The reported intraoperative hypotension and successfully treated for an anaphylactic reaction per the device ifu is most likely device related due to a suspected polymer leak.As identified in fs-0012, the root cause for most polymer leaks is a material weakness adjacent to the polymer fill channel which may become compromised during pressurization with liquid polymer.Since this device remains implanted, endologix is unable to conduct product evaluation to conclusively ascertain root cause.However, based on the investigation associated with this fsn, this is the most likely cause associated with this event.On 06aug2018, endologix issued a field safety notice (fs-0009) regarding the risk of polymer leak events with the ovation ix aortic body stent graft.On 06may2020, endologix issued a follow-up safety update (fs-0012) reaffirming treatment recommendations for patients who experience a polymer leak during implantation and providing updated information on the current rate of polymer leaks, the rate of clinical harms and root cause.The manufacturing lot evaluation confirmed all devices met specifications prior to release.No additional investigation of this reported event is planned however if additional information pertinent to the incident is obtained, a follow-up report will be submitted.Endologix will continue to monitor this complaint and similar complaints in the event further investigation is needed.H6: result code: remove code 3233; h6: conclusion code: remove code 11.
|