MDT POWERED SURGICAL SOLUTIONS MIDAS REX®LEGEND FOOTED ATTACHMENT; MOTOR, DRILL, PNEUMATIC Back to Search Results
Model Number AF01
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/20/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Report confirmed.Evaluation determined that a portion of the foot of the attachment was perforated.It was also noted the bearings were corroded and tool could not be inserted into the attachment.Previous investigation performed under (b)(4) determined that the likely causes are debris in the collet and improper insertion of the tool.The user manual contains the following warnings ¿do not use a legend attachment if any part of the attachment appears to be bent, loose, missing or damaged.Do not use excessive pressure, such as bending or prying, on attachments or dissecting tools.This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff." in addition, ¿insert dissecting tool into motor collet with a slight rotational motion.A tactile and audible click is observed indicating that the dissecting tool is fully seated.¿ we will continue to track and trend this complaint type.Initial reporter: (b)(6).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Repair request initiated for device with the report of damaged attachment.It was reported that there was no patient involvement.Re pair request escalated to a product event based on reason for return.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX®LEGEND FOOTED ATTACHMENT
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
MDR Report Key8716011
Report Number1625507-2019-00047
Device Sequence Number11393761
Product Code HBB
UDI-Device Identifier00721902481840
UDI-Public00721902481840
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAF01
Device Catalogue NumberAF01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2019
Initial Report FDA Received Date06/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-