| Catalog Number |
530.715 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device evaluation: the actual device was returned for evaluation.During repair, it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to premature wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported by (b)(6) that during service and evaluation, it was determined that the battery reciprocator device made an unexpected noise.It was further determined that the device had a fluid ¿liquid¿ leak.It was noted in the service order that the device had an undetermined malfunction and it made an abnormal noise.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2019.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
| |
|