| Catalog Number |
CAT5 |
| Medical Device Problem Codes |
Break (1069); Device Damaged Prior to Use (2284)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
| |
|
Event or Problem Description
|
|
During preparation for a thrombectomy procedure, the physician found that an indigo system cat5 aspiration catheter (cat5) was broken into two pieces upon removal from the packaging.The damage to the cat5 was found prior to use.Therefore, the cat5 was not used in the procedure.The procedure was completed using a new cat5.
|
| |
|
Additional Manufacturer Narrative
|
|
Results: the indigo system cat5 aspiration catheter (cat5) was fractured approximately 77.0 cm from the hub.The device was ovalized approximately 127.5 cm from the hub.Conclusions: evaluation of the returned cat5 confirmed a fracture.If the device is forcefully manipulated at extreme angles during removal from its packaging hoop, damage such as a fracture may occur.Further evaluation revealed a distal ovalization.This damage was likely incidental to the reported complaint.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|