| Model Number |
24950KLQ |
| Medical Device Problem Codes |
Device Emits Odor (1425); Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the remote monitor had a strong burning odor and was hot.It was also reported that a brown burnt spot was noted under the remote monitor.The caller was instructed to immediately unplug and place the remote monitor on a non-flammable surface.The remote monitor is expected to be replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: model#: 24950klq, serial/lot#: (b)(4): the remote monitor was returned, and analysis revealed that there was no complaint failure found; found error codes (b)(4) in failure analysis (fa) log.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The remote monitor was returned and replaced.
|
| |
|