Catalog Number 1013469-200
Medical Device Problem Codes Leak/Splash (1354); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/30/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4).Evaluation summary: visual and functional inspection was performed on the returned device.The reported leak was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The investigation determined that the leak appears to be related to a potential product quality issue.It should be noted that the armada 18 otw instruction for use (ifu) states: do not use if the device is damaged.In this case, continuing to use the device did not contribute to the leak.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion with mild calcification in the superficial femoral artery (sfa).During preparation of the 5.5 x 200 mm armada 18 balloon catheter, some air bubbles were observed, but the armada 18 was advanced in the vessel.When the balloon was pressurized to nominal pressure a leak was observed at the hub.The balloon catheter was removed and exchanged for a new armada 18 to complete the procedure.There was no clinically significant delay in the procedure and no adverse patient effect.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8718109
Report Number2024168-2019-04910
Device Sequence Number16740960
Product Code LIT
UDI-Device Identifier08717648208355
UDI-Public08717648208355
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K151317
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 06/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Catalogue Number1013469-200
Device Lot Number6051241
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2019
Initial Report FDA Received Date06/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age61 YR