| Model Number |
DBP-125MICRO145 |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Failure analysis conclusion: the saline line was not returned for analysis.Fluid was run through the oad and observed flowing out the distal end of the saline sheath as intended.When tested the oad functioned as intended.At the conclusion of the failure analysis investigation, the flushing event could not be not confirmed through analysis.(b)(4).
|
| |
|
Event or Problem Description
|
|
Difficulty was encountered during attempted flushing of the diamondback peripheral orbital atherectomy device (oad).Resistance was encountered during attempts to flush air from the saline tubing.Despite troubleshooting, air could not be purged from the tubing, and a second oad was used to successfully continue the procedure.The patient was well following the procedure.
|
| |
|