• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM; CORONARY ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM; CORONARY ATHERECTOMY DEVICE Back to Search Results
Model Number DBP-125MICRO145
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/21/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Failure analysis conclusion: the saline line was not returned for analysis.Fluid was run through the oad and observed flowing out the distal end of the saline sheath as intended.When tested the oad functioned as intended.At the conclusion of the failure analysis investigation, the flushing event could not be not confirmed through analysis.(b)(4).
 
Event or Problem Description
Difficulty was encountered during attempted flushing of the diamondback peripheral orbital atherectomy device (oad).Resistance was encountered during attempts to flush air from the saline tubing.Despite troubleshooting, air could not be purged from the tubing, and a second oad was used to successfully continue the procedure.The patient was well following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM
Common Device Name
CORONARY ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
MDR Report Key8718285
Report Number3004742232-2019-00158
Device Sequence Number2091412
Product Code MCX
UDI-Device Identifier10852528005282
UDI-Public(01)10852528005282(17)210331(10)261538
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P130005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2021
Device Model NumberDBP-125MICRO145
Device Catalogue NumberDBP-125MICRO145
Device Lot Number261538
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2019
Initial Report FDA Received Date06/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-