| Model Number |
S702 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/17/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This device has not been returned; therefore, a technical analysis cannot be conducted.Without a returned device it is not possible to definitively confirm how the device may have contributed to the complaint incident.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
| |
|
Event or Problem Description
|
|
It was reported that the device battery had allegedly prematurely depleted.The device is pending explant to resolve the event at an unspecified future date.The patient was stable with no adverse consequences and no further information is available.
|
| |
|
Event or Problem Description
|
|
It was reported that the device battery had allegedly prematurely depleted.The device was explanted and replaced to resolve the event.The patient was stable with no adverse consequences.
|
| |
|
Event or Problem Description
|
|
It was reported that the device battery had allegedly prematurely depleted.The device was explanted and replaced to resolve the event.The patient was stable with no adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.A high current drain was detected.Detailed analysis confirmed that a high current drain was the result of a compromised capacitor, which resulted in the observed premature battery depletion.Investigation has determined that the high current drain is the result of compromised low voltage capacitors, which is caused by the presence of excess hydrogen from a liner component within the device.
|
| |
|