| Catalog Number |
71333325 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
05/30/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during surgery the liner not locked properly in to the shell.They used another liner that also not locked.They removed reflection shell 54 and used r3 shell 56 and liner.30 minutes of delay due to this incident.A backup device was available.
|
| |
|
Additional Manufacturer Narrative
|
|
Reviewing the information received by the manufacturer has identified that this event should be re-evaluated for mdr reporting.The information states that the patient suffered bone loss during the reflection shell removal and not while using the insert in the patient, therefore, it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated complaint devices were not returned.The clinical/medical team concluded, without the requested clinical information, a thorough medical investigation cannot be rendered.Should any additional clinical information be provided this complaint will be re-assessed.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed parts revealed no prior complaints for the listed batches.Without the actual device involved our investigation cannot proceed.If the device or new information is received in the future, the complaints can be re-opened.No further actions are being taken at this time.
|
| |
|
Search Alerts/Recalls
|