| Model Number |
1207-36 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/06/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
It was reported that the device exhibited premature battery depletion.The device was explanted and replaced.The patient was in a stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Interrogation of the device revealed the device was at eri when received.A longevity calculation was performed and found the battery depletion was normal based on the device usage.Based on this information, there was no evidence of premature depletion.
|
| |
|