| Model Number |
0015790 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleges while loading a patient into the ambulance the drop frame allegedly bent when the foot end of the cot was lifted.The operators then chose to have the patient transferred to another stretcher to complete the transfer.No injuries or delay in treatment were reported as a result of the incident.
|
| |
|
Event or Problem Description
|
|
Complainant alleges while loading a patient into the ambulance the drop frame allegedly bent when the foot end of the cot was lifted.The operators then chose to have the patient transferred to another stretcher to complete the transfer.No injuries or delay in treatment were reported as a result of the incident.
|
| |
|
Additional Manufacturer Narrative
|
|
The cot was evaluated by an authorized field technician at the complainants site.A visual and functional evaluation wasconducted.The technician found the stretcher to be in proper working order however; the technician found the customer had incorrectly insalled the straps of an accessory to the drop frame.This allowed the drop frame to bend when pressure was applied to the accessory.The ifu provides clear instruction for the proper installation of the accessory.
|
| |
|