| Model Number |
7885 |
| Medical Device Problem Codes |
Fracture (1260); Difficult to Advance (2920); Material Deformation (2976); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/31/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that tip fracture occurred.An impulse catheter was selected for use.During the procedure outside the patient, it was observed that the tip was fractured and the device would not advance.The procedure was competed with another of the same device.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction.Initially reported device code fracture 1260 based on additional information received on 28june2019, the device code has been updated to material deformation 2976.G1: mfr site address 2: no.20905 int.A.Col., cd.Industrial, 22444.
|
| |
|
Event or Problem Description
|
|
It was reported that tip fracture occurred.An impulse catheter was selected for use.During the procedure outside the patient, it was observed that the tip was fractured and the device would not advance.The procedure was completed with another of the same device.There were no patient complications reported.It was further reported that the tip was observed to be kinked and the patient's status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer; analysis of the hub, shaft and tip included microscopic and visual inspection.Analysis revealed kinks in the shaft located 2.4cm, 5.5cm and 66cm from the tip of the device.Inspection of the rest of the device found no other damage.G1: mfr site address 2: no.20905 int.A.Col., cd.Industrial, 22444.
|
| |
|
Event or Problem Description
|
|
It was reported that tip fracture occurred.An impulse catheter was selected for use.During the procedure outside the patient, it was observed that the tip was fractured and the device would not advance.The procedure was completed with another of the same device.There were no patient complications reported.It was further reported that the tip was observed to be kinked and the patient's status was stable.
|
| |
|