OBERDORF SYNTHES PRODUKTIONS GMBH 4.5MM LCP® MEDIAL PROXIMAL TIBIA PLATE 10H/RIGHT 214MM; PLATE, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
239.992 |
| Medical Device Problem Codes |
Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Impaired Healing (2378)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Complainant part is not expected to be returned for manufacturer review/investigation.No facility contact available; initial reporter is company representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported on an unknown date, a revision surgery was scheduled due to a possible refracture and or non-union.The construct originally implanted was implanted off-label.The revision surgery was scheduled for (b)(6) 2019 and will be performed by another surgeon.Original implant date was (b)(6) 2019.It is unknown if the re-scheduled surgery has occurred.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part 239.990, lot 7678882: part manufacture date: may 12, 2014.Manufacturing location: elmira.A review of the device history record revealed no complaint related anomalies.The device history record shows this lot of 4.5mm lcp medial proximal tibia plate 10h/right 214mm product was processed through the normal manufacturing and inspection operations with no rework nor nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.A review of the raw material device history record revealed this lot met all specifications with no nonconformance noted.This raw material lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.H3, h6: a product investigation was completed: the plate was received broken into two pieces with the transverse fracture located at the second combi-hole.No other issues were identified with the returned plate.Dimensional inspection showed the relevant dimensions were conforming.The relevant drawings, reflecting the manufactured and current revision, were reviewed.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.No definitive root cause could be determined.It is possible that condition was due to early excessive strain by patient and/or unintended forces.During this investigation no product design or manufacturing issues were observed that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
During investigation of the returned devices it was noted the plate as broken.Concomitant devices: cortex screw (part 214.838, lot 9202997, 9271477, l888929, quantity 3); cortex screw (part 214.840, lot l663008, quantity 1); locking screw (part 212.212, lot 9201970, quantity 1); screw (part/lot unknown, quantity 1).
|
| |
|