Catalog Number 7205305 |
Device Problem
Material Discolored (1170)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/03/2019 |
Event Type
malfunction
|
Event Description
|
It was reported that during a knee procedure, it was noticed that the tip of the blade was black.There was no delay or patient injuries but, it is unknown how the procedure was finished since no backup device was available to complete it.Attempts were made to retrieve further details about the event but the complainant does not have more information.
|
|
Manufacturer Narrative
|
One 7205305 3.5mm full radius disposable blade used for treatment, was not returned for evaluation.A customer photo was provided.The device appeared in good condition.Silicone is used in during manufacturing along with a silver plating process.The silver plate surface may become discolored.Visual inspection indicated evidence of silicone and silver tarnish discoloration.To produce a device with less friction, silicone is used for additional lubrication between inner and outer blades.This may encourage subsequent tarnish of the silver.It has been determined that the excess silicone located at the bottom of the blade tip, or the appearance of discoloration will not impact the patient or end user.¿surillium is the proprietary, silver coating that helps prevent shedding.It improves lubricity and allows for reduced frictional contact between the stainless steel inner and outer blades.It also discolors when exposed to certain elements.The red substance that's occasionally found on our blades and peaks curiosity of our customers is actually a silicone lubricant from the manufacturing process.Testing has confirmed that the silicone lubricant is biocompatible, non-toxic, non-irritating and non-sensitizing.¿ the silicone and silver improves performance of the device and has been determined to have no adverse effect on patient safety.This observation is monitored through surveillance.Complaint history lot review found one other complaint for this lot.Product met specifications upon release to distribution.No further investigation is warranted at this time.
|
|
Search Alerts/Recalls
|
|
|