| Model Number |
MMT-723RNAL |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/16/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer¿s husband reported via phone call that the insulin pump had a blank display.The customer¿s blood glucose level was unknown at the time of the incident.Customer states the contacts on the battery cap are neither missing nor damaged, battery compartment was neither damaged nor corroded and the spring was neither damaged nor corroded.Trouble shooting was performed for blank display.Customer was advised to clean battery cap contacts with a dry cotton swab, if no iso propyl alcohol is available to use an alcohol wipe or dry cotton swab.Advised to allow at least one minute for the battery contacts to dry and to insert new aa battery.Display did not return after pump restart.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The customer was advised the pump will be replaced as per troubleshoot.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump monitored for several days and no blank display noted.Insulin pump received with normal operating current.Insulin pump passed self test.Insulin pump passed off no power test.
|
| |
|