MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 20 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 20 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 20
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/29/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was not returned to physio-control for evaluation.The cause of the reported issue could not be determined.Device not evaluated by manufacturer.
 
Event Description
The customer contacted physio-control to report that they were using this device to do a cardioversion on an infant patient and the unit was set to 4 joules; however, the device's printout showed that 200 joules was delivered to the patient.There were no reports of any adverse effects to the patient as a result of the reported issue.The customer informed physio-control that no additional patient information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 20 DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key8727603
MDR Text Key148971795
Report Number0003015876-2019-01080
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873813744
UDI-Public00883873813744
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 06/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number99402-000016
Was Device Available for Evaluation? No
Date Manufacturer Received05/30/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/25/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-