• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT PSL HA CLUSTER 58MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT PSL HA CLUSTER 58MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 542-11-58G
Medical Device Problem Codes Crack (1135); Fracture (1260)
Health Effect - Clinical Codes Injury (2348); Reaction (2414); Device Embedded In Tissue or Plaque (3165)
Date of Event 05/29/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the product history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported that the patient's right hip was revised after patient complaint of pain.Intra- operatively, black tissue and ceramic fragments were noted, and no ceramic was inside of the alumina insert.The shell, 2 screws, alumina liner and alumina head were revised.Rep provided a primary operative report and implant sheet, pre-revision x-rays, and explant pictures.
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture involving a trident shell reported.Conclusions: it was reported that patient was revised due to pain.The review of the provided medical records by clinical consultant state that: the x-rays include an ap and lateral the right hip.They demonstrate an uncemented right tha with 2 screws in the acetabular shell, with the acetabular and femoral components well fixed and nominally positioned.An alumina head is reduced into the acetabular component but is eccentrically migrated superiorly and several radio-dense fragments are noted adjacent to the joint.This is consistent with the description of a fragmented, alumina, acetabular insert".Need revision operative reports, office/clinical reports, serial x-rays, etc.Based on the provided information the product reported in this investigation did not contribute to the event and is therefore considered concomitant.As this case refer to crack/fracture between liner and head so shell not contributed to failure.No further investigation for this event is possible at this time as no devices and / or insufficient information was received by stryker orthopaedics.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported that the patient's right hip was revised after patient complaint of pain.Intra- operatively, black tissue and ceramic fragments were noted, and no ceramic was inside of the alumina insert.The shell, 2 screws, alumina liner and alumina head were revised.Rep provided a primary operative report and implant sheet, pre-revision x-rays, and explant pictures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT PSL HA CLUSTER 58MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8727913
Report Number0002249697-2019-02355
Device Sequence Number1685737
Product Code LPH
UDI-Device Identifier07613327037319
UDI-Public07613327037319
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2009
Device Catalogue Number542-11-58G
Device Lot Number17147801
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/29/2019
Supplement Date Received by Manufacturer08/26/2019
Initial Report FDA Received Date06/24/2019
Supplement Report FDA Received Date09/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
-
-