DEPUY SYNTHES PRODUCTS LLC ELECTRIC PEN DRIVE 60,000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.010 |
Device Problem
Unintended Movement (3026)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/01/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Udi: (b)(4).The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined the device run in the locked position.It was also observed that the device failed pretest for check function and direction of rotation.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that during service and evaluation, it was determined that the electric pen drive device ran in the locked position.It was further determined that the device failed pretest for check function and direction of rotation.It was noted in the service order that the device had an unspecified malfunction.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|