• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC ELECTRIC PEN DRIVE 60,000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC ELECTRIC PEN DRIVE 60,000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 05.001.010
Device Problem Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2019
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined the device run in the locked position.It was also observed that the device failed pretest for check function and direction of rotation.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Event Description
It was reported from (b)(6) that during service and evaluation, it was determined that the electric pen drive device ran in the locked position.It was further determined that the device failed pretest for check function and direction of rotation.It was noted in the service order that the device had an unspecified malfunction.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRIC PEN DRIVE 60,000 RPM
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
kara ditty-bovard
eimattstrasse 3
oberdorf 4436
SZ   4436
6103142063
MDR Report Key8728141
MDR Text Key149085748
Report Number8030965-2019-65601
Device Sequence Number1
Product Code HWE
UDI-Device Identifier07611819167711
UDI-Public7611819167711
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K043310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 06/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2019
Initial Date Manufacturer Received 06/11/2019
Initial Date FDA Received06/24/2019
Date Device Manufactured04/11/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-