| Model Number |
500AHCT |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/30/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The srt replaced the arterial bpm.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
The service repair technician (srt) reported that during routine testing of the device at the service center, the arterial blood parameter monitor (bpm) failed the high potential (hi-pot) test.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected block: e1 the reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|