BAXTER HEALTHCARE - MOUNTAIN HOME; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number ASKU
Device Problems Disconnection (1171); Improper Flow or Infusion (2954); Protective Measures Problem (3015)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/26/2019
Event Type  malfunction  
Manufacturer Narrative
An alarm indicative of a potential malfunction of the disposable cassette was reported.The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.However, a disconnection was reported between the supply bag and the supply line, which is known to cause this alarm.The cause of the disconnection could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a homechoice device experienced an unspecified alarm.The patient was connected at the time of the alarm.This occurred during dwell three of peritoneal dialysis therapy.During the troubleshooting, air was observed in the supply line.It was further reported that the supply bag had a loose connection and became disconnected from the supply line.Renal therapy services (rts) advised the patient to start all over with new supplies and reviewed proper procedures per the user manual.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NI
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
mountain home AR
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8728238
MDR Text Key149241280
Report Number1416980-2019-03413
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 06/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/26/2019
Initial Date FDA Received06/24/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age91 YR
-
-