• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #6; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #6; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6021-0637
Medical Device Problem Codes Degraded (1153); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Injury (2348)
Date of Event 05/29/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.A review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported that the patient's hip was revised due to trunnion damage.Update: "the head disassociated from the stem.Shell was not revised.Converted the liner to mdm.".
 
Event or Problem Description
It was reported that the patient's hip was revised due to trunnion damage.Update: "the head disassociated from the stem.Shell was not revised.Converted the liner to mdm.".
 
Additional Manufacturer Narrative
An event regarding fretting and disassociation involving a lfit v40 cocr head that was mated with an accolade stem was reported.The event of fretting was confirmed by the provided photographs and the event of disassociation was not confirmed.Method & results: -device evaluation and results: the reported device was not returned however photographs were provided for review.The photographs shows a recently explanted stem with blood and tissue on it.Pencilling of the proximal tip of the trunnion is present on the stem which is consistent with fretting having occurred.-clinician review: a review of the provided medical records and/or x-rays by a clinical consultant indicated: "date of implant (b)(6) 2010, date of explant 5/29/19.Primary tha and revision tha implant labels note revision to rest.Modular stem and mdm bearing.1 undated x-ray ap of right hip demonstrates an uncemented primary tha, reduced with suspicion of head/trunnion junction disruption.No operative reports or findings at revision surgery available.Insufficient data to create a report." -device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusion: the reported accolade stem was mated with a lfit v40 cocr head which has been identified to be within scope of nc and capa.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.No further investigation is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE PLUS TMZF HIP STEM #6
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8728263
Report Number0002249697-2019-02356
Device Sequence Number13900130
Product Code LPH
UDI-Device Identifier04546540510716
UDI-Public04546540510716
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2015
Device Catalogue Number6021-0637
Device Lot Number33505901
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/29/2019
Supplement Date Received by Manufacturer07/16/2019
Initial Report FDA Received Date06/24/2019
Supplement Report FDA Received Date08/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age86 YR
-
-