| Model Number |
R SERIES |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the device and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed an "ecg disabled" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the malfunction was observed during review of the device's history log.The device was put through extensive testing without duplicating the malfunction.The analog board was replaced as a precaution.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|