|
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime test, excessive no delivery test and delivery accuracy test.The stop current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.After completing the rewind test, installed in the reservoir compartment.Device recognized the water filled reservoir that was installed in the reservoir compartment.Then the unit was programmed with a 2.0 unit fill cannula delivery.The unit delivered the 2.0 units properly with water exiting the infusion set.No rewind anomaly or prime/fill anomaly noted during testing.Device uploaded properly using care link.Device had minor scratched display window.(b)(4).
|