• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC INSERTER; DRIVER, PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC INSERTER; DRIVER, PROSTHESIS Back to Search Results
Model Number 273511300
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4), unknown.The device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, during routine incoming inspection of a loaner set, it was observed that the inserter was broken.There was no patient involvement.This complaint involves one (1) device.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Udi: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Visual examination of the returned found the proximal end of the handle is broken.There are light scratches along the handle and the shaft that are consistent with wear.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.Although no definitive root cause could be determined, it is possible that the device encountered unintended forces during use or storage.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSERTER
Common Device Name
DRIVER, PROSTHESIS
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
MDR Report Key8730432
Report Number1526439-2019-51786
Device Sequence Number13958613
Product Code HWR
UDI-Device Identifier10705034190587
UDI-Public(01)10705034190587
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number273511300
Device Catalogue Number273511300
Device Lot NumberAG2066617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2019
Initial Date Received by Manufacturer 06/04/2019
Supplement Date Received by Manufacturer07/29/2019
Initial Report FDA Received Date06/25/2019
Supplement Report FDA Received Date07/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-