| Model Number |
TN,DXH 900 HEMATOLOGY SYSTEM WITH FLOOR STAND |
| Medical Device Problem Code |
Installation-Related Problem (2965)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
05/21/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
While installing the instrument the fse injured his hand.The flap behind the floor stand (compressor side) of the instrument was jammed and opened jerkily.The fse was able to complete the instrument install process.Bec internal identifier - (b)(4).
|
| |
|
Event or Problem Description
|
|
In the process of installing the dxh 900 hematology instrument at the customer site, the field service engineer (fse) injured the index finger on his left hand when a flap in the rear area of the instrument opened jerkily.The fse was treated by a doctor.The hand was x-rayed and the finger was found to be bruised and fractured.The injury did not require stitches but the finger was stabilized with a rail/splint.
|
| |
|