• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY SEE H10; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY SEE H10; INTRAOCULAR LENS Back to Search Results
Model Number CC4204A
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/28/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Collamer ultraviolet-absorbing posterior chamber single piece foldable intraocular lens.(b)(4).
 
Event or Problem Description
The reporter indicated that a cc4204a, +23.00 diopter, intraocular lens was removed with no patient injury and replaced with another staar lens due to posterior vault.The reporter stated " equator to equator diameter is to small, caused lens to posterior vault".
 
Additional Manufacturer Narrative
Device evaluation: lens was returned in liquid in a lens vial.Visual inspection found the haptic torn and the lens was returned in three pieces.Claim#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEE H10
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key8731917
Report Number2023826-2019-01115
Device Sequence Number2271352
Product Code HQL
UDI-Device Identifier00841542102452
UDI-Public00841542102452
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P990013
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2019
Device Model NumberCC4204A
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2019
Initial Date Received by Manufacturer 05/29/2019
Supplement Date Received by Manufacturer08/30/2019
Initial Report FDA Received Date06/25/2019
Supplement Report FDA Received Date09/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-