| Model Number |
CC4204A |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
05/28/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Collamer ultraviolet-absorbing posterior chamber single piece foldable intraocular lens.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a cc4204a, +23.00 diopter, intraocular lens was removed with no patient injury and replaced with another staar lens due to posterior vault.The reporter stated " equator to equator diameter is to small, caused lens to posterior vault".
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: lens was returned in liquid in a lens vial.Visual inspection found the haptic torn and the lens was returned in three pieces.Claim#: (b)(4).
|
| |
|