MENTOR TEXAS MENTOR SILTEX CONTOUR PROFILE HIGH; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3542715 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
05/24/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6)-year-old caucasian female patient underwent a breast augmentation primary with a mentor siltex contour profile high 650cc and experienced deflation on the right breast implant.As a result, the patient will undergo explantation on (b)(6) 2019.
|
| |
|
Additional Manufacturer Narrative
|
|
On 7/25/2019, the mentor failure analysis lab has received the device for evaluation.On 8/1/2019, the analysis results showed a tear at the union between patch and shell of the implant.Additionally, leak testing revealed a second tear noted in the posterior aspect and measuring approximately 1.5 cm.Microscopic examination of the edges of both ruptures revealed parallel striations on one of the tears.No other anomalies were observed.The striations are consistent with markings made by a sharp instrument perforating silicone material.As stated in the product insert data sheet, do not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps or scissors to contact the device during the implantation or other surgical procedures as this can result in rupture.The manufacturing records were reviewed, and the manufacturing criteria were met prior to the release of this lot.It should be noted that as part of our quality process, each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Based on the information reported and/or the product investigation, there is no evidence that the issue is related with the manufacturing process.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 9/19/2019, it was reported to mentor that the replacement breast implant was mentor siltex contour profile high 650cc filled to 725cc.The patient had a thickened capsule along the right inflammatory scar ¿ submitted for pathology.Manufacturer¿s reference number: (b)(4).
|
| |
|