| Model Number |
7130 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/19/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of an emerge balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple minor kinks along the hypotube.There is blood present in the hub, inflation lumen and balloon.The balloon was loosely folded prior to inflation testing.The presence of dried blood made the device unable to be inflated so it was soaked in water bath for several days.The device was then inflated, and a leak was noticed coming from the balloon.Microscopic examination revealed a pinhole in the balloon 5mm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6) 2019.It was reported that shaft kink occurred.The 95% stenosed target lesion was located in the left anterior descending artery (lad).A 1.50mm x 8mm emerge balloon catheter was used for dilatation but the delivery shaft was kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed balloon pinhole.
|
| |
|