Model Number 305901
Medical Device Problem Codes Break (1069); Difficult to Remove (1528)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/29/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: product id :353101, serial# unknown, product type: external neurostimulator.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received by a trial patient regarding an external neurostimulator (ens) for urge incontinence.Patient stated she had such a headache and has been nauseated.Patient stated her she is not voiding as much as well.Patient advised to contact her clinician.Stylet wouldn¿t pull out from pne lead.There was none external or environmental factors to have occurred.Physician tried to pull it out for several minutes before it eventually broke.After it broke, the stylet retracted back into the lead.We retested the lead with good results, so physician decided to leave as was after determining it wouldn¿t affect hook up to the extension.There was no surgical intervention planned or done.The issue was resolved at the time of this event.There were no further complications reported at this time.
 
Event or Problem Description
Additional information received from a manufacturer representative (rep) stated that the cause was not determined.The stylet will not be returned as it was pulled out and then discarded during normal wire pull at the physician's office.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
MDR Report Key8733451
Report Number3007566237-2019-01384
Device Sequence Number11422366
Product Code EZW
UDI-Device Identifier00643169721166
UDI-Public00643169721166
Combination Product (Y/N)N
PMA/510(K) Number
P970004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date12/04/2020
Device Model Number305901
Device Catalogue Number305901
Device Lot Number60168413
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2019
Supplement Date Received by Manufacturer06/26/2019
Initial Report FDA Received Date06/26/2019
Supplement Report FDA Received Date06/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10...."
Patient Age85 YR
-
-