| Catalog Number |
04P52-01 |
| Medical Device Problem Codes |
Failure to Calibrate (2440); High Test Results (2457)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/20/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation into this issue determined that there is a high possibility that control and calibrator materials had been mixed during the fill process and that hba1c calibrator lot 54582uq02 may contain a quantity of high control material instead of calibrator level 2 (l2).A product recall letter was issued to all architect customers who have received the architect hba1c calibrator lot 54582uq02.This lot may generate an active calibration curve, a shift to qc and patient results may occur.The letter instructs the customer to discontinue use of the suspected lots and destroy any remaining inventory.
|
| |
|
Event or Problem Description
|
|
The customer received error code 1353 calibration failure, calibration factor out of range and controls out of range high while using the architect hba1c calibrator, list 04p52-01, lot 54582uq02.No discrepant patient results generated.No impact to patient management was reported.
|
| |
|