| Model Number |
24690 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/30/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that foreign material was found in the device.A 3.5mmx30mmx135cm (4f) sterling balloon catheter was selected for use.However, during unpacking, a hair was found on the holder of the device.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that foreign material was found in the device.A 3.5mmx30mmx135cm (4f) sterling balloon catheter was selected for use.However, during unpacking, a hair was found on the holder of the device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reported city (b)(6).Device evaluated by mfr.: returned product consisted of a sterling balloon catheter inside the packaging.The product package, pouch, and device were visually inspected.The box and pouch were both open.The pouch was opened along the poly-tyvek seal.The bsc seal was still intact.The device was still in its carrier tubing.There was foreign matter consisting of a hair on the carrier tube inside the pouch.Inspection of the remainder of the packaging and device presented no other damage or irregularities.
|
| |
|