| Catalog Number |
SECX-10-40-135 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/19/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the physician used a protege rx stent to treat a severely calcified, moderately tortuous, calcified lesion in the mid left internal carotid-common carotid artery.There was no damage noted to the packaging and no issues noted when removing the device from the tray.The device was prepped per ifu without issue.A 6fr non-medtronic sheath and a medtronic guardwire were used for the procedure along with guardwire as embolic protection.The device did not pass through a previously deployed stent there was no resistance encountered nor excessive force used during delivery of the device to the lesion.Approx 3months post procedure, on (b)(6) 2019, the patient returned for a follow up visit and the physician observed that the stent was inverted and was seen as ¿heart shaped¿.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: the protégé rx stent shape appeared normal after surgery.The physician had planned to perform graft replacement for the thoracic aortic aneurysm.The patient is currently under dapt.Patient is being monitored due to dapt.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|