ARTHREX, INC. TIGERSTICK, #2 TIGERWIRE; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Model Number TIGERSTICK, #2 TIGERWIRE
Medical Device Problem Codes Patient-Device Incompatibility (2682); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Erythema (1840); Unspecified Infection (1930); Inflammation (1932); Irritation (1941); Loss of Range of Motion (2032); Reaction (2414); Fluid Discharge (2686)
Date of Event 04/26/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The device remains in patient.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that the patient underwent an acl reconstruction procedure on (b)(6) 2019 where the following devices were implanted: ar-1588rt-j, lot: 10259067; ar-1897s, lot: 10224925; ar-4030c-10, lot: 10198681; ar-7209t, lot: 12863; jrf speedgraft id: 183528-015 (not an arthrex product).This case was completed with no issue.The patient was seen for his first post-op visit on (b)(6) 2019 and per the pa: ¿skin intact; port and graft incisions clean, dry and intact with sutures.No erythema or signs of infection.Mild redness secondary to suture reaction.¿ the patient was seen for a 2nd follow-up visit and the surgeon reported: ¿35ml of serosanguinous fluid aspirated from the right knee after alcohol prep, no hint of turbidity.¿ this was done in response to increased dwelling and disability for three days.Surgeon reports: ¿not clinically a dvt; infection a concern, but also, clinically, does not look like it.¿ patient was placed on bactrimds by mouth twice daily for seven days.The fluid was sent for cell count.Per the report, the fluid was cloudy with total nucleated cell count at 6800 and neutrophils at 75.The patient was seen for his third post-op follow up visit on (b)(6) 2019.The patient was still having right knee effusions with limited rom and pain.The following was reported: ¿no redness or drainage but there is an effusion and 25cc of slightly turbid yellow fluid aspirated.¿ aspirate fluid sent for cell count and per the report, the fluid was cloudy with total nucleated cell count at 31800 and neutrophils at 85.The surgeon ordered the patient to continue bactrim.On (b)(6) 2019, the patient underwent an arthroscopic washout and a limited synovectomy to the right knee for possible post-op septic.The patient received 2g iv antibiotics and was washed out with triple antibiotic irrigation.On (b)(6) 2019 the surgeon stated that the patient¿s c-reactive protein is double normal indicating a non-specific inflammation.The rep spoke with the surgeon on (b)(6) and he stated that the patient has completed a 2 week course of iv antibiotics and that he is treating the patient as if he has an infection but the cultures have come back negative.As of (b)(6) 2019, the cell count for the aspirate fluid is 2000.The surgeon stated that the patient is possible having a reaction to the graft.Rep requested follow-up visit reports from (b)(6) 2019 to (b)(6) 2019 and upon receipt of reports, the patient had a follow-up visit where he underwent a s/p surgery on (b)(6) 2019.The patient is on iv antibiotics at home.The surgeon indicates that the culture done during the (b)(6) 2019 follow-up showed no growth.Surgeon states in his follow up notes: ¿the dx of infection after his acl reconstruction is not a certainty but something is severely irritating his knee and we are erring on the side of caution.The plan is 3 weeks of iv vanco and cipro with weekly lab tests.¿ additional information provided (b)(6) 2019: the op notes and md office are unaware if the original acl repair was done with arthrex products.Patient has previous bilateral acl reconstruction in 1994 and 1996.In between follow-ups, the patient was admitted to the hospital on (b)(6) 2019 and discharged (b)(6) 2019.Patient followed by home health for iv abx via picc line.Iv abx therapy completed (b)(6) 2019.As of last visit with the surgeon ((b)(6) 2019), the infection resolved.Per notes "resolved post op infection though we never grew any germs." the surgery was for a washout only.There were no arthrex products removed and there is no plan to have any arthrex products removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIGERSTICK, #2 TIGERWIRE
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key8734922
Report Number1220246-2019-01168
Device Sequence Number11406172
Product Code GAT
UDI-Device Identifier00888867041394
UDI-Public00888867041394
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K122374
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Model NumberTIGERSTICK, #2 TIGERWIRE
Device Catalogue NumberAR-7209T
Device Lot Number12863
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2019
Initial Report FDA Received Date06/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age51 YR
-
-