• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY Back to Search Results
Model Number G34785
Device Problem Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/28/2019
Event Type  malfunction  
Manufacturer Narrative
Pma/510(k) # k142688.(b)(4).1 unit of lot c1520192 of echo-hd-3-20-c was returned opened in its original packaging.The device related to this occurrence underwent a laboratory evaluation.A proximal kink was observed below the sheath extender.Prior to distribution, all echo-hd-3-20-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at (b)(4).A review of the manufacturing records for echo-hd-3-20-c of lot number c1520192 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1520192.The notes section of the instructions for use, which accompanies this device instructs the user to inspect the device prior to use : "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".There is no evidence to suggest that the customer did not follow the instructions for use.The failure of needle kinked/bent was observed in the laboratory.A definitive root cause for the customer complaint could not be determined as circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to the device being used in a flexed or twisted position as indicated in the additional information causing the issue with needle advancement.Complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event Description
The outer sheath surface creased and the needle cannot be advanced and retracted during procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE
Type of Device
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
Manufacturer (Section G)
COOK IRELAND LTD
o halloran road
national technology park
limerick
Manufacturer Contact
heather ryan
o halloran road
national technology park
limerick 
061334440
MDR Report Key8735225
MDR Text Key150146633
Report Number3001845648-2019-00286
Device Sequence Number1
Product Code FCG
UDI-Device Identifier00827002347854
UDI-Public(01)00827002347854(17)210712(10)C1520192
Combination Product (y/n)N
Reporter Country CodeCN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Remedial Action Modification/Adjustment
Type of Report Initial
Report Date 05/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/12/2021
Device Model NumberG34785
Device Catalogue NumberECHO-HD-3-20-C
Device Lot NumberC1520192
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/25/2019
Event Location Hospital
Date Manufacturer Received05/30/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/12/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
Patient Weight60
-
-