• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. DOUBLECUT, 4.0MM X 13CM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. DOUBLECUT, 4.0MM X 13CM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Model Number DOUBLECUT, 4.0MM X 13CM
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/07/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that during a shoulder arthroscopy procedure, the surgeon was doing a decompression and the oval burr, ar-8400obe was producing metal flakes in the joint.No further information.Additional information requested.Additional information provided 6/12/2019: the shavings looked like glitter in the joint and were probably not retrieved.The case was completed with a new shaver.Additional information provided 6/13/2019: the account has two blades from the case.They also used an ar-8400dc, lot: 10274224.In addition to the ar-8400obe and they aren¿t 100% sure which one produced the metal flakes, so they saved both.
 
Additional Manufacturer Narrative
The complaint could not be confirmed.Visual evaluation revealed no surface damage to the inner diameter of the ar-8400dc outer hub, or any damage on the corresponding location at the distal end of the inner tuber that would indicate a site of metal debris production.Material analysis showed the returned ar-8400dc met all as-received specifications.No problem found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOUBLECUT, 4.0MM X 13CM
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key8735392
Report Number1220246-2019-01176
Device Sequence Number19400827
Product Code GFA
UDI-Device Identifier00888867043565
UDI-Public00888867043565
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model NumberDOUBLECUT, 4.0MM X 13CM
Device Catalogue NumberAR-8400DC
Device Lot Number10274224
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 06/07/2019
Supplement Date Received by Manufacturer06/07/2019
Initial Report FDA Received Date06/26/2019
Supplement Report FDA Received Date07/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-