MAUDE Adverse Event Report: DEPUY MITEK LLC US 5.5 HEALIX ADVANCE PEEK W/OC; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 5.5 HEALIX ADVANCE PEEK W/OC; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Catalog Number 222305
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/27/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Event or Problem Description
It was reported by the affiliate via phone that during a right rotator cuff repair procedure the 5.5 healix advance peek anchor with orthocord suture loop was broken and could not fix the anchor.Another device was used to complete the surgery.There were no adverse consequences to the patient.The surgery time was not delayed.No additional information could be provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received and inspected.A visual inspection was performed to determine if the device had any gross visual defects that may contribute to the complaint condition.However,it is observed that healix advance peek anchor with orthocord suture loop was broken and small broken fragment was observed.Anchor breakages are typically associated with off axis insertion, levering during insertion,hard bone quality or using incorrect instrumentation for preparing bone hole.The possible root cause for the reported failure could be user mishandling.However the definitive root cause cannot be determined.No nonconformances were identified for this part number, lot number combination per qlik query executed on 5/31/2019.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 HEALIX ADVANCE PEEK W/OC
Common Device Name
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8735705
Report Number1221934-2019-57441
Device Sequence Number11406151
Product Code HWC
UDI-Device Identifier10886705021413
UDI-Public10886705021413
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K120449
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2019
Device Catalogue Number222305
Device Lot Number3919778
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2019
Initial Date Received by Manufacturer 05/31/2019
Supplement Date Received by Manufacturer08/16/2019
Initial Report FDA Received Date06/26/2019
Supplement Report FDA Received Date08/19/2019
Patient Sequence Number1
-
-