• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CRE RX; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CRE RX; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00558930
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/03/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre rx biliary dilatation balloon was used in the papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, while physician was performing sphincteroplasty, a small hole opened in the side of the balloon and water started to stream out causing the balloon to deflate.The procedure was completed with a different device.There have been no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: problem code 1504 captures the reportable event of balloon pinhole.Block h10: investigation results visual examination of the returned complaint device revealed the balloon did not show visual defects and looked normal.No damage found on the catheter of the device.Functional analysis was performed, and the balloon was inflated without a problem; however, balloon would not hold pressure due to a pinhole found over the ro marker.This failure is likely due to factors or conditions related to the procedure that could have affected its performance and its intended purpose such as; the manner the device was handled, and/or the interaction with the scope during the procedure.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre rx biliary dilatation balloon was used in the papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, while physician was performing sphincteroplasty, a small hole opened in the side of the balloon and water started to stream out causing the balloon to deflate.The procedure was completed with a different device.There have been no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE RX
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8737051
Report Number3005099803-2019-03238
Device Sequence Number13929441
Product Code FGE
UDI-Device Identifier08714729965480
UDI-Public08714729965480
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K172520
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/04/2022
Device Model NumberM00558930
Device Catalogue Number45767
Device Lot Number0023290456
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2019
Initial Date Received by Manufacturer 06/03/2019
Supplement Date Received by Manufacturer07/10/2019
Initial Report FDA Received Date06/26/2019
Supplement Report FDA Received Date08/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-