(b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional h3: it was reported performance pull off suture needle an empty opened box and twenty-two representative samples were returned for evaluation.The complaint sample was not received.During the visual inspection of thirteen samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, or damaged were observed.A functional test was performed and the pull force were above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the conditions of the samples received, no performance pull off suture needle was found, and the tested samples met the finished goods requirements.Representative samples returned to support investigation of 6 device issues captured in reports 2210968-2019-83325, 2210968-2019-83326, 2210968-2019-83327, 2210968-2019-83328, 2210968-2019-83329, and 2210968-2019-83331.Analysis results have been included in follow-up reports for each.
|