• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number ATG80146
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 06/06/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lot number of the device was provided.The device history records are currently under review.The return of the sample is pending.The investigation is currently underway.(expiry date: 12/2021).
 
Event or Problem Description
It was reported that during an angioplasty procedure in the superior vena cava, the pta balloon allegedly detached and migrated to the pulmonary artery.It was further reported that a sternotomy was performed to remove the detached balloon.The patient was reported stable post procedure.
 
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was returned for evaluation.A visual inspection found the balloon to be inverted and completely detached from the catheter shaft.Therefore, the investigation is confirmed for the reported balloon detachment.The definitive root cause for the reported balloon detachment could not be determined based upon available information.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the superior vena cava, the pta balloon allegedly detached and migrated to the pulmonary artery.It was further reported that a sternotomy was performed to remove the detached balloon.The patient was reported stable post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8737594
Report Number2020394-2019-01257
Device Sequence Number2573282
Product Code LIT
UDI-Device Identifier00801741060779
UDI-Public(01)00801741060779
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K181323
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberATG80146
Device Catalogue NumberATG80146
Device Lot Number93YC0158
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2019
Initial Date Received by Manufacturer 06/06/2019
Supplement Date Received by Manufacturer11/04/2019
Initial Report FDA Received Date06/26/2019
Supplement Report FDA Received Date11/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
-
-