| Model Number |
ZXT375 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Visual Disturbances (2140); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided, but the best estimate date is (b)(6) 2019.Device evaluation: product testing could not be performed because the product was not returned.The reported complaint was not confirmed.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.The search revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.(b)(4).All pertinent information available to johnson and johnson surgical vision has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a tecnis symfony toric multifocal intraocular lens (model zxt375 +15.5 diopter) was explanted from patient¿s left eye in a secondary surgical procedure because the patient was over refracted, and experiencing specific visual issues.At explant, the incision was enlarged, and sutures were placed.Another lens, a model zlb00 +15.0 diopter was used as the replacement lens for the patient.Patient was reported stable at discharge.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|
Search Alerts/Recalls
|