• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723NAB
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/30/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer reported via phone call that the insulin pump was not functioning and had motor error.Customer reports they were unable to describe the possible damage to the drive support cap the customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The device will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the displacement test.Device received with motor error alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform prime/a33 test, excessive no delivery test and occlusion test due to motor error alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAB
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8738721
Report Number2032227-2019-22015
Device Sequence Number16653998
Product Code OYC
UDI-Device Identifier00613994743893
UDI-Public(01)00613994743893
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723NAB
Device Catalogue NumberMMT-723NAB
Device Lot NumberA3723NABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2019
Initial Date Received by Manufacturer 06/20/2019
Supplement Date Received by Manufacturer09/28/2019
Initial Report FDA Received Date06/27/2019
Supplement Report FDA Received Date10/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age46 YR
Patient Weight205
-
-